Low-potential amperometric determination of hydrogen peroxide with a carbon paste electrode modified with nanostructured cryptomelane-type manganese oxides

Yuehe Lin,Xiaoli Cui,Liyu Li
DOI: https://doi.org/10.1016/j.elecom.2004.12.005
IF: 5.443
2005-01-01
Electrochemistry Communications
Abstract:Nanostructured cryptomelane-type manganese oxides were synthesized, characterized, and evaluated for chemical sensing. Cryptomelane-type manganese oxides are nanofibrous crystals with sub-nanometer open tunnels that provide a unique property for sensing applications. Carbon paste electrodes (CPEs), modified with the nanostructured cryptomelane-type manganese oxides, were investigated for amperometric detection of hydrogen peroxide. With an operating potential of +0.3 V versus Ag/AgCl, H2O2 produces catalytic oxidation currents at the modified CPE, which can be exploited for quantitative determinations. The amperometric signals are linearly proportional to H2O2 concentration in the range 1.0×10−4–6.9×10−4 M with a correlation coefficient of 0.995 (n=7). At a signal-to-noise ratio of 3, a detection limit of 2 μM can be observed for the CPE modified with 5.5 wt% cryptomelane-type manganese oxides. In addition, the sensor has a good stability and reproducibility. The construction and renewal are simple and inexpensive. A possible response mechanism was proposed and discussed. The significant electrocatalytic activity of the modified CPE may result from the nanostructure of cryptomelane-type manganese oxides.
What problem does this paper attempt to address?