Characterization of Nanogold-based Quartz Crystal Microbalance DNA Biosensor Array Using Electrochemical Impedance Spectroscopy

XIA Han,WANG Feng,HUANG Qing,FU Wei-ling,HUANG Jun-fu
2010-01-01
Abstract:OBJECTIVE To Characterize nanogold-based quartz crystal microbalance(QCM) DNA biosensor array using electrochemical impedance spectroscopy(EIS). METHODS Faradaic impedance spectra measurements were performed as following: the bare electrode, after thiolated probe modification, after 6-mercaptohexanol modification, after the hybridization with target DNA and after the signal amplification with nanoparticles. In addition, the frequency shifts were also recorded. RESULTS The frequency shift caused by the conjugation of Au nanoparticles to target DNA (127.5±6.8Hz) was significantly larger than that caused by the hybridization, the blank control and the negative control (P0.01). The interfacial electron-transfer resistances significantly increased after each step (P0.01). CONCLUSIONS The EIS is proved to be a non-damage and useful technique for characterizing the multilayer films on the quartz crystal. The results well demonstrate the fabrication and detection process on the quartz crystal.
What problem does this paper attempt to address?