Estimating Subglottal Pressure Via Airflow Redirection

Christopher D. Baggott,Alexander K. Yuen,Matthew R. Hoffman,Liang Zhou,Jack J. Jiang
DOI: https://doi.org/10.1097/mlg.0b013e318063e89e
2007-01-01
Abstract:Subglottal pressure (SGP) is a valuable parameter in the research and clinical assessment of laryngeal function. The lungs serve as a constant pressure source during sustained phonation, and that pressure SGP, can be used to determine the efficiency with which the larynx converts aerodynamic power (SGP x glottal airflow) has been shown to reliably reflect vocal health. However, current SGP measurement techniques are hesitantly used because of either an invasive nature or the requirement of intensive patient training. This study tests a novel device that has been designed to noninvasively estimate SGP through mechanical airflow redirection, producing a numeric output on completion of the trial which lasts only 135 ms, which is predicted to limit the effect of confounding laryngeal reflexes that may occur during the trials. Additionally, the airflow redirection in to the retension device allows for the pneumatic in-trial comparison of the estimated SGP with the pressure achieved by the patient, providing a numeric output to the clinician on completion.
What problem does this paper attempt to address?