Application Of Multi-Optical Path Length Modeling In The Quantitative Analysis Of Human Whole Blood

Li Gang,Yuliang Liu,Lin Ling,Wang Yan
DOI: https://doi.org/10.3321/j.issn:0253-3820.2007.10.022
IF: 1.193
2007-01-01
Chinese Journal of Analytical Chemistry
Abstract:In order to raise the predicting resolution of the calibration model, the quantitative analysis of six blood compositions by method of multi-optical path length is conducted. Blood NIRS is measured by the optical path length tunable system and the component s information is tested by the automatic biochemical analyzer. The partial least square method (PLS2) is used to model and predict the blood compositions. Due to the nonlinear characteristic relationship between the blood spectra and the concentration of the blood composition, the equivalent absorption coefficient of the blood sample of multi-optical path length is different. The NIR spectra of the human blood of multi-optical path length (from 0.20 unit to 1.25 mm) at same wavelength are irrelevancy. Here, taking nonlinear characteristic as a method actively, no modeling using one optical path length, but multi-optical path length, the concentrations of blood composition are predicted. It is revealed by the research that the quantitative analysis of whole blood by the method of multi-optical path length modeling method can raise the resolution of the blood calibration model.