Electrochemical sensor for multiplex biomarkers detection.

Fang Wei,Prabhudas Patel,Wei Liao,Kishore Chaudhry,Lei Zhang,Martha Arellano-Garcia,Shen Hu,David Elashoff,Hui Zhou,Shilin Shukla,Franky Shah,Chih-Ming Ho,David T Wong
DOI: https://doi.org/10.1158/1078-0432.CCR-09-0050
IF: 13.801
2009-01-01
Clinical Cancer Research
Abstract:PURPOSE:Multiplexing assay of biomarkers at the point-of-care is an elusive goal for molecular diagnostics. EXPERIMENTAL DESIGN:Here, we report an electrochemical (EC) sensor for oral cancer detection based on the simultaneous detection of two salivary biomarkers: interleukin (IL)-8 mRNA and IL-8 protein. RESULTS:Under the multiplexing mode, the limit of detection of salivary IL-8 mRNA reaches to 3.9 fM and 7.4 pg/mL for IL-8 protein in saliva. Multiplex assay of these 2 biomarkers directly from 28 cancer and 28 matched control saliva samples shows significant difference between the two groups. From the receiver operating characteristic analysis, the EC sensor yields around 90% sensitivity and specificity for both IL-8 mRNA and IL-8 protein, which are very close to the data measured by traditional assays (ELISA and PCR) with the same group of saliva. Combined IL-8 mRNA and protein show better AUC compared with single biomarker. CONCLUSIONS:We show, for the first time, concurrently multiplexing detection of salivary mRNA and protein biomarkers using point-of-care EC sensor.
What problem does this paper attempt to address?