A Dynamic Micro-Ct Scanner Based on a Carbon Nanotube Field Emission X-Ray Source

G. Cao,Y. Z. Lee,R. Peng,Z. Liu,R. Rajaram,X. Calderon-Colon,L. An,P. Wang,T. Phan,S. Sultana,D. S. Lalush,J. P. Lu,O. Zhou
DOI: https://doi.org/10.1088/0031-9155/54/8/005
2009-01-01
Abstract:Current commercial micro-CT scanners have the capability of imaging objects ex vivo with high spatial resolution, but performing in vivo micro-CT on free-breathing small animals is still challenging because their physiological motions are non-periodic and much faster than those of humans. In this paper, we present a prototype physiologically gated micro-computed tomography ( micro-CT) scanner based on a carbon nanotube field emission micro-focus x-ray source. The novel x-ray source allows x-ray pulses and imaging sequences to be readily synchronized and gated to non-periodic physiological signals from small animals. The system performance is evaluated using phantoms and sacrificed and anesthetized mice. Prospective respiratory-gated micro-CT images of anesthetized free-breathing mice were collected using this scanner at 50 ms temporal resolution and 6.2 lp mm(-1) at 10% system MTF. The high spatial and temporal resolutions of the micro-CT scanner make it well suited for high-resolution imaging of free-breathing small animals.