Accurate Measurement Of Neodymium Isotopic Composition Using Neptune Multiple Collector Inductively Coupled Plasma Mass Spectrometry

Yang Yue-Heng,Zhang Hong-Fu,Xie Lie-Wen,Wu Fu-Yuan
DOI: https://doi.org/10.3321/j.issn:0253-3820.2007.01.013
IF: 1.193
2007-01-01
Chinese Journal of Analytical Chemistry
Abstract:This paper reports the measurement of the neodymium isotopic composition by neptune multiple collector inductively coupled plasma mass spectrometry (MC-ICP-MS) over last two years. For the cerium concomitant with neodymium in the chemical separation, there is no significant influence on the neodymium analysis. As for the sample containing small amount of samarium (Sm/Nd < 0. 04), direct calibration for isobaric interference and mass discrimination by the exponential law can be obtained by assuming that samarium mass discrimination is the same as that of neodymium. Geological samples after traditional chemical separation were measured by neptune MC-ICP-MS and thermal ionization mass spectrometry (TIMS), respectively. The results show that Neptune MC-ICP-MS can precisely measure neodymium isotopic composition as the TIMS does, even more effective and less time-consuming than the TIMS Method.