Determination concentration of melatonin in human saliva by HPLC-MS/MS

WEN Bing-ting,XIONG Xin,LIU Na,SHEN Yang,ZHAO Rong-sheng
DOI: https://doi.org/10.3969/j.issn.1001-6821.2010.04.015
2010-01-01
Abstract:Objective To establish a HPLC-MS/MS method for the determination of melatonin in human saliva. Methods Saliva samples and internal standard (itopride) were extracted by dichloromethane after alkalization. The analysis was conducted using a Sunfire C18 column with gradient elution and the column temperature is 35 ℃. The mobile phase consisted of methanol and water with 0.1% formic acid at flow rate of 0.3 mL·min-1. Analysis in mass was operated by the electrospray ionization. Results LLOQ of the method is 5.48 pg·mL-1(S/N>10). The standard curve is linear in the range of 5.48 ~250.00 pg·mL-1(γ2=0.9994). LLOD is 2 pg·mL-1(S/N>3). The RSDs of inter-day and intra-day are all less than 15%. The extraction recovery is closed to 60%. Conclusion The method is fast, precise and sensitive, and suitable for the determination of melatonin in human saliva.