Simutaneous Determination of Lactulose, Mannitol and Lactose in Urine with Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) Method

胡强,徐红兵,李水军,刘罡一,陆品相,余琛
DOI: https://doi.org/10.3969/j.issn.1005-8982.2008.13.005
2008-01-01
Abstract:[Objective] To establish a LC-MS/MS method for the determination of lactulose, mannitol and lactose in human urine. [Method] Urine samples were simply diluted with water and acetonitrile. Chromatographic separation was performed on Phenomenex Luna NH2 (100×2.0 mm, 5 μm) column using acetonitrile-water (80:20, v/v) as mobile phase at flow rate of 0.3 mL/min. The analysis were protonated in the positive electrospray ionization (ESI) interface and detected in multiple reaction monitoring (MRM) mode with the transitions from the precursor to the production of m/z 181.1/88.9 for mannitol, m/z 341.3/160.9 for lactulose and lactose and m/z 353.4/166.9 for internal standard (IS). [Results] The retention time of mannitol, lactulose, lactose and IS were found at 2.9, 4.1, 4.7 and 4.7 min. Calibration curve was linear in the range from 0.1 to 40 μg/mL with a correlation coefficient above 0.9980. The lower limit of quantitation was 0.1 μg/mL for mannitol and lactulose, 0.2 μg/mL for lactose. The intra- and inter-day precision (CV%) were 0.60%~4.31% and 2.55%~7.71%, the intra- and inter-day accuracy were 86.4%~107.6% and 95.9%~102.9%, the recovery was 84.7%~105.5%. [Conclusion] The LC-MS/MS method for the quantitative determination of mannitol, lactulose and lactose in human urine was simple, fast, sensitive and specific, which can be used for mannitol and lactulose determination for intestinal permeability assessment in clinic applications.