A Loading Device for Intervertebral Disc Tissue Engineering

Lihui Fu,Chunqiu Zhang,Baoshan Xu,Dong Xin,Jiang Li
DOI: https://doi.org/10.1007/978-3-642-18134-4_90
2011-01-01
Abstract:In recent years, the new technologies related with tissue engineering improving continuously bring the hope for the complete regeneration of the morphology and physiological function of the degenerated intervertebral disc. According to the biomechanics of intervertebral disc, from the mechanical characteristic of human activities, we have developed a dual-frequency compression-torsion loading device for intervertebral disc tissue engineering. The device consists of three systems. The dual-frequency compression part contains two systems which produces low-frequency high-amplitude compression (frequency 0-3Hz, amplitude 0-3mm) and high-frequency low-amplitude (frequency 10-90Hz, amplitude 0-40μm). Rotary system provides torsion loading to the culture by back and forth rotation controlled by step moter. The maximum rotation angle is set smaller than 30°. The design goal of the loading device is to achieve the mechanical condition of dual-frequency compression- torsion loading as close as possible to that of in vivo condition of the native disc tissue. And also is used to study the mechanobiology of intervertebral disc.