The Detection of Pulse Wave Condaction Time and Its Clinical Application in Wave Intensity Technique

肖沪生,银浩强,徐智章,张爱宏,徐芳,彭欣,靳炜,陆盈,邬云燕,朱蓓菁,高东雯,王奇
2008-01-01
Abstract:Objective To explore the feasibility of detecting pulse wave conduction time and its application in evaluating arterial elas-ticity in clinic by the technique of wave intensity(WI).Methods The pulse wave conduction time of carotid and right brachial artery of 66 healthy adults were detected by Prosound α10 color Doppler ultrasonography manufactured by company of ALOKA.Results The pulse wave conduction time from the beginning of aorta to left common carotid artery,right common carotid artery and right brachial artery respectively were(25.17±9.53)ms、(26.64±8.43)ms and(73.76±12.15)ms.There were all statistical significances in the pulse wave conduction time differ-ence in left and right common carotid arteries(P0.05),common carotid arteries of both sides and right brachial artery(P0.01).The body height were positively correlated with the pulse wave conduction time from the beginning of aorta to the left(r=0.360,P0.01) or right(r=0.397,P0.01) common carotid artery and right brachial artery(r=0.417,P0.01).Conclusions The farther the measuring point is away from the heart,and the higher the body is,the longer the pulse wave conduction time is.The pulse wave velocity,which is one of classic indexes in evaluating arterial elasticity,can be calculated according to the distance from the measuring between points and pulse wave conduction time.
What problem does this paper attempt to address?