Control of endotracheal cuff pressures using finger estimation and manometer measurement

He Ru,He Jie,Huang Yan,Tao Yanling
DOI: https://doi.org/10.3870/hlxzz.2012.17.004
2012-01-01
Abstract:Objective To explore an effective method for measurement and control of endotracheal cuff pressure(Pcuff).Methods One hundred consecutive mechanically ventilated patients were randomized to a control group(n=46) and an observation group(n=54).Initial target Pcuff(20-30 cmH2O) of both groups was set with a manometer,and then Pcuff of the control group was monitored and adjusted with a manometer according to current recommendations(three times a day).For the observation group,after initial measurement with manometers,Pcuff was additionally monitored by using finger estimation: 5 nurses were trained to master measurement of Pcuff with finger estimation and were asked to adjust Pcuff every 4 hours with this method.Results Compared with the control group,the ratio of Pcuff between 20 and 30 cmH2O in the observation group was significantly higher,and incidence rates of complications and cuff leakage were signi-ficantly reduced(P<0.05,P<0.01);Nurses′ adherence to monitoring of Pcuff in the observation group was significantly increased(P<0.01).Conclusion Monitoring of Pcuff with finger estimation and manometer measurement is safe and effective,which can improve nurses′ compliance with measurement and reduce incidence of complications.
What problem does this paper attempt to address?