Determination of 25 Organophosphorus Pesticide Residues in Tea by Isotope-Labelled Internal Standard Quantitation and High Performance Liquid Chromatography Tandem Mass Spectrometry

王敏,韩芳,张蕾,金葆康,郑平,胡艳云,宋伟,吕亚宁
DOI: https://doi.org/10.13595/j.cnki.issn1000-0720.2013.0107
2013-01-01
Abstract:A sensitive and effective isotope-labelled internal standard method for rapid determination of 25 organophosphorus pesticides in tea has been developed.The sample was extracted by mixture of ethyl acetate and n-hexane,purified using Carb-NH2 solid phase extraction(SPE) cartridges and high performance liquid chromatography tandem mass spectrometric(LC-MS/MS) detection.For accurate quantitative analysis,three isotope-labelled organophosphorus were compared,and stable Diazinon-d10 was chosen as the internal standard at last.The average recoveries of each pesticide were 70%~108% at three spiked concentration levels,and the relative standard deviations(RSDs) were between 2.5% and 11.3%,the limits of quantitation(LOQs) were all below 8.0 μg/kg,and good linear relationships were observed with the correlation coefficients r2 0.997 for all analytes.The recovery,repeatability of method can meet analysis requirements of organophosphorus residues in tea.
What problem does this paper attempt to address?