[Compare NRT Value of Straight Electrode and Contour Electrode in Nucleus 24 Cochlear Implant System].

Yong Tao,Yun Zheng,Kai Wang,Zhaoli Meng,Ke Xu
DOI: https://doi.org/10.13201/j.issn.1001-1781.2012.01.006
2012-01-01
Abstract:OBJECTIVE:To learn the neural response of spiral ganglion cell to electrical stimulus by comparing the neural response telemetry(NRT) value between straight electrode and contour electrode in Nucleus 24 cochlear implant system.METHOD:Fourty-six patients with Nucleus 24 cochlear implants were paired into two groups: 23 with contour electrode and 23 with straight electrode respectively. NRT value of each group were tested and compared.RESULT:The NRT value of straight electrode fell in the range of 160 microV to 170 microV and contour electrode 150 microV to 160 microV. The NRT were going up higher a little bit from electrode twenty to electrode one.CONCLUSION:The contour electrode requires less power consumption than straight electrode in Nucleus 24 cochlear implant system.