Abstract:T-wave alternans(TWA) is one of significant and important diagnostic parameters of heart disease patients.However,quantitative analysis of TWA is far from easy.This article adopted the integral average of fixed length method instead of the traditional single reference mode to improve the performance of quantitative analysis of TWA.The data revealed that anti-interference capabilities of time domain were significantly improved.Moreover,we utilized the synthesized linear model to simulate artificial TWA signals.Based on variability and heterogeneity of physiological signals in practical application,we revised the David S.group proposed estimation formula of TWA magnitude.The results demonstrated that TWA magnitude could be quantitative measurement and the absolute error could be controlled within ± 5 μV when SNR of signal is greater than 10 dB.
What problem does this paper attempt to address?