Microfluidic device as a new platform for immunofluorescent detection of viruses.

Wen-Tso Liu,Liang Zhu,Qi-Wei Qin,Qing Zhang,Hanhua Feng,Simon Ang
DOI: https://doi.org/10.1039/b509086e
IF: 7.517
2005-01-01
Lab on a Chip
Abstract:A bead-based microfluidic device was developed and demonstrated to achieve rapid and sensitive enzyme-linked immunosorbent assay ( ELISA) with quantum dots as the labeling fluorophore for virus detection. In comparison to standard ELISA performed on the same virus, the minimal detectable concentration of the target virus was improved from 360 to 22 ng mL(-1), the detection time was shortened from >3.25 h to <30 min, and the amount of antibody consumed was reduced by a factor of 14.3.
What problem does this paper attempt to address?