Application of MTI photoscreener in amblyopia of infants

丁娟,赵堪兴,郑曰忠
DOI: https://doi.org/10.3760/cma.j.issn.2095-0160.2008.04.014
2008-01-01
Abstract:Objective Analyzing visual damage form photos has been applied to large scale epidemiological survey for preschool childhood.But different referral criteria will given different results.Present study was to investigate the feasibility of Medical Technology and Innovation(MTI)Photoscreener as a new referral criteria in the screening of infants amblyopia.Methods244 eyes from 122 1-2-year old subjects were examined in Tianjin Eye Hospital with MTI Photoscreener and Cycloplegic retinoscopy.The outcome for screening of amblyopia with two methods was compared.ResultsThe sensitivity of screening of amblyopia with MTI Photoscreener was 92.16% and specificity was 65% with the positive forecasting value 93.07% and negative forecasting value 61.90%.The highest sensitive factor by MTI Photoscreener was myopia(100%)and the lowest one was strabismus(60.00%).The highest specific factor by MTI Photoscreener was anisometropia(100%),and the lowest one was astigmatism(94.57%).In the screening of different types of strabismus,the highest sensitivity of MTI Photoscreener was vertical strabismus(75%)and the lowest one was exptropia(55.55%),and the sensitivity for esotropia was 66.67%.ConclusionPhotoscreening is of a high sensitivity in the screening of amblyopia in infants.It offers a reliable method for epidemiology investigation.
What problem does this paper attempt to address?