Organophosphorus hydrolase multilayer modified microcantilevers for organophosphorus detection.

Chandana Karnati,Hongwei Du,Hai-Feng Ji,Xiaohe Xu,Yuri Lvov,Ashok Mulchandani,Priti Mulchandani,Wilfred Chen
DOI: https://doi.org/10.1016/j.bios.2006.10.027
IF: 12.545
2007-01-01
Biosensors and Bioelectronics
Abstract:We report a biosensor based on organophosphorus hydrolase (OPH) multilayer modified microcantilever (MCL) for detection of organophosphorus compounds (OPs). The assay is based on substrate-dependent bending of the OPH functionalized MCLs. The cantilever bending amplitude at equilibrium was a function of the concentration of paraoxon with the dynamic range extending from 10−7 to 10−3M. The lower detection limit of approximately 10−7M for paraoxon was an order of magnitude better than the OPH-based potentiometric and optical biosensors based on pH modulation. There was a good intra-sensor and an acceptable inter-sensor reproducibility as evidenced by the standard errors of 5% and 15%, respectively. OPs measured using this technique included parathion and diisopropyl fluorophosphate (DFP) in the order of paraoxon>DFP>parathion. The conformational change of the OPH was most likely the main origin of MCL bending.
What problem does this paper attempt to address?