Percutaneous screw fixation of acetabular fractures under fluorscopy-based computerized navigation

LUO Cong-feng,GAO Hong,ZHOU Kai-hua,HU Cheng-fang,ZENG Bing-fang
DOI: https://doi.org/10.3969/j.issn.1674-134x.2008.02.015
2008-01-01
Abstract:Objective To evaluate the preliminary clinical outcome of the fluoroscopy-based computerized navigation system used in percutaneous screw fixation on acetabular fractures.Methods From May 2005 to August 2007,33 patients with 34 acetabular fractures were treated with percutaneous screw fixation under the guidance of a fluoro-navigation system.According to AO classification,21 acetabular fractures were of Grade A3,8 of Grade B1,2 of Grade B2 and 3 of Grade C1.The average interval from injury to surgery was 5.6 days(2-8 days).Two patients underwent open reduction with mini incision.Results 43 hollow screws were inserted.On the average,every hollow screw needed 24.6 minutes to be inserted and 28.4 seconds of X-ray exposure.The images of fluoro-navigation were compared with those of the real X-rays both intra-and post-operatively.The mean deviated distance was 1.96 mm and the mean deviated angle was 2.45 degrees in the postoperative verification.The intraoperative blood loss was less,screw deviation out of the fracture occurred in 1 case during the operation.1 patient sustained a transient femoral nerve palsy but recovered two months after operation,which was attributed to the limited open reduction.No infection and implant failure occurred.Evaluation by modified d'Aubigne and Postel clinical scoring,excellent result was in 25 cases,good in 5 and fair in 3,the excellent and good rate was 91%.Conclusion Percutaneous screwing is an effective early management on the treatment of acetabular fractures.Assistance with fluoro-navigation,it makes more precise,safe,time-saving and less X-ray exposure to do screw fixation on the acetabular fractures.
What problem does this paper attempt to address?