Surface Characteristics and Biocompatibility of Three Metals for Cardiovascular Stents

WU Feng,MENG Gaili,DONG Binghui,LIU Baoguo,ZHOU Yimin,FU Tao
DOI: https://doi.org/10.3969/j.issn.1672-6278.2012.02.006
2012-01-01
Journal of Biomedical Engineering Research
Abstract:Surface characteristics,hemocompatibility and endothelium of 316L stainless steel,NiTi alloy and TLM titanium alloy were investigated for the selection of cardiovascular stent material.316L stainless steel had good hemocompatibility and endothelium,but weak corrosion resistance.NiTi alloy exhibited overall good properties.TLM alloy,which is composed of biocompatible elements,had good corrosion resistance,but its high surface roughness and weak hydrophilicity brought down its hemocompatibility and endothelium capacity.For cardiovascular stent application,NiTi alloy has the best biocompatibility among the three materials.
What problem does this paper attempt to address?