Abstract:Objective:To evaluate the possibility and accuracy of real-time three-dimensional echocardiography(RT-3DE) in the assessment of the mechanical prosthetic valve apparatus's configuration and function in patients with prosthetic valves replacement,and to explore its superiority and the clinical value.Methods:RT-3DE and TTE were used to observe the structure,neighbor relationship of artificial valve and its bracket,and evaluate hemodynamic changes in fifty patients experienced the mitral valve(MV) replacement or/ and aortic valve(AV) replacement separately.Results:①92.6%(25/27) patients with MV replacement,56.5%(13/23) patients with AV replacement,MV in 73.3%(11/15) patients with MV and AV replacement could be successfully observed by gray-scale RT-3DE.②color RT-3DE could be used to assess the valve regurgitation and the leakage around the mechanical prosthetic valve.The outcome of RT-3DE method and TTE method is rather consistent,the Kappa value of judgment by the two methods was 0.72 in the valve regurgitation,and 0.74 in the perivalvular leakage,respectively(P<0.001),and the Spearman correlation value was 0.72 and 0.89,respectively(P<0.001).Conclusion:Mechanical posthetic valve apparatus's structure,motion and the flow dynamics could be accurately detected by RT-3DE.Compared with transthoracic echocardiography(TTE),RT-3DE can provide more additional information of anatomic structure and hemodynamics.