Determination of Cardiac Troponin I by Anodic Stripping Voltammetry at SBA-15 Modified Carbon Paste Electrode

HS Guo,NY He,SX Ge,D Yang,JN Zhang
DOI: https://doi.org/10.1016/s0167-2991(05)80275-7
2005-01-01
Abstract:A novel electrochemical immunoassay for cardiac troponin I (cTnl) combining the concepts of the dual monoclonal antibody "sandwich" principle, the silver enhancement on the nano-gold particle, and the SBA-15 modified carbon paste electrode (SBA-MCPE) is described. Four main steps were carried out to obtain the analytical signal, i.e. electrode preparation, immunoreaction, silver enhancement, and anodic stripping voltammetric detection. A linear relationship between the anodic stripping peak current and concentration of cTnI from 0.5 to 5.0 ng/ml and a limit of detection of 0.2 ng/ml were obtained. The established method was tested by determining cTnI in acute myocardial infarction (AMI) samples using enzyme-linked immunoadsorbent assay (ELISA) for comparison analysis, and good results were obtained.