Mechanically Experimental Study on a Capsule Robot in Vitro

Zhang Wei,Huang Ping
DOI: https://doi.org/10.3969/j.issn.0254-0150.2007.06.005
2007-01-01
Abstract:A new method for driving a micro capsule robot moving in vitro was presented.A permanent magnet array produces a rotating magnetic field.The capsule robot is with a pair of magnetic bars and a spiral surface.A tester was built and some samples were fabricated.Several types of viscosity silicon oil simulating body fluid were confected in experiments.Some experiments were carried out in the silicon oil and in a small intestine of a pig in vitro separately.The experimental results show that in the central area of the magnet array,the strength and rotational speed of magnetic field are adjustable serially,but the rotational direction is contrary to that of the permanent magnet.When the diameter of the array is fixed,the strength of the rotational magnetic field is proportional to the number of the magnets.When the robot is in the viscosity fluid,the diameter of the robot,the speed of rotation and the viscosity of the fluid decide the strength of the outer rotational magnetic field and the resistance torque together.Only with a suitable diameter can a spiral-type micro capsule robot move through the pig intestine in vitro.To the robot samples moving through the pig intestine freely,the resistance force is under 0.2 N and the resistance torque is lower than 2.6 mN·m.Therefore,if the torque is up to 3.2 to 4.5 mN·m loaded on the robot samples by the outer rotational magnetic field,it is enough to drive the robot samples to revolve and move back-and-forth reliably.
What problem does this paper attempt to address?