RP- HPLC Determination of Content in Doxorubicin Hydrochloride Magnetic Liposomes

ZHAO Wen,ZHANG Hong,LU Ting-li,CHEN Tao
DOI: https://doi.org/10.16155/j.0254-1793.2010.02.035
2010-01-01
Abstract:Objective:To establish an RP-HPLC method for determine content in doxorubicin hydrochloride mag-netic liposomea. Methods: The separation was performed with Agilent HC-C_(18) column (250 mm × 4.6 mm, 5μm), the mobile phase consists of a mixture of methanol and acetic acid/sodium acetate solution (pH 3.4) (7:3)at a flow rate of 1.0 mL·min~(-1), with UV detection at 233 nm and ambient temperature. Results: The concentration of doxorubicin hydrochloride had good linear correlation in the range of 2.0-50.0 μg·mL~(-1)(r=0.9994, n=7),the average recovery was 97.1%. Conclusions:This method is simple, rapid, accurate, sensitive and reliable for de-termination of content and entrapment efficiency in doxorubicin hydrochloride magnetic liposomes.
What problem does this paper attempt to address?