New Technique for Mouse Oocyte Injection Via a Modified Holding Pipette

Q. F. Lyu,L. Deng,S. G. Xue,S. F. Cao,X. Y. Liu,W. Jin,L. Q. Wu,Y. P. Kuang
DOI: https://doi.org/10.1016/j.rbmo.2010.07.004
IF: 4.567
2010-01-01
Reproductive BioMedicine Online
Abstract:To improve mouse oocyte survival from intracytoplasmic sperm injection, the sharp tip of the injection pipette has been modified to have a flat end. Here, for the same goal but for a more convenient manipulation, a sharp injection pipette was kept whereas the holding pipette was modified to have a trumpet-shaped opening, which allows deeper injection into the oocyte as it is held. Mouse oocyte injection with mouse and human spermatozoa was performed at 37°C. For the injection of mouse oocyte with mouse sperm head, a significantly higher survival rate (83%) was achieved by utilizing the modified holding pipette than the conventional one (21%; P<0.001) and the fertilization rates were normal and comparable for both methods (82% versus 81%). A superior survival rate (82%) and acceptable normal fertilization rate (71%) were also achieved by utilizing the modified holding pipette for interspecies ICSI (injecting mouse oocyte with human spermatozoon). Taken together, by utilizing a holding pipette with a trumpet-shaped opening, acceptable rates of mouse oocyte survival and fertilization can be achieved using a sharp injection pipette under conditions usual for human oocyte injection.