In Negative Ion Mode MALDI-TOF MS Analysis of Amino Acids Using Carbon Nanotubes As Matrix

CHEN Jin-an,HUANG Hui-ying,HUANG He-qing
DOI: https://doi.org/10.3321/j.issn:0438-0479.2007.06.029
2007-01-01
Abstract:Multiwalled carbon nanotubes were used as the matrix for analysis of small molecules by MALDI-TOF MS,and it was observed tube able to eliminate interference of matrix background ions and suitable for analysis of small molecules(0.8 ku).Here,we demonstrated this method through analyze amino acids and amino acid derivatives in positive reflection mode.Mass spectra showed that the analytes were ionized as sodium adduct+ and potassium adduct+,metal ions are the main chemical adduct and+ peaks of the analytes are not observed in the mass spectra.Previously reported method mainly focused on positive reflection mode.In this work we focused on the development of this new matrix for MALDI to analyze amino acids in negative reflection mode.It was observed from the mass spectra that there was only one intense peak,which was corresponding to-,and under optimal conditions,little or no background peaks appeared in the low mass region,and the intensity and S/N of peaks was enhanced.This method can be utilized in the rapid and sensitive analysis of amino acids,and expand the applied fields of MALDI-TOF mass spectrometry.
What problem does this paper attempt to address?