Online Detecting Techniques for Atmospheric Pollutants in Submarine Compartments

张骧,侯可勇,陈少杰,王卫国,李海洋
DOI: https://doi.org/10.3969/j.issn.1673-3185.2012.05.019
2012-01-01
Abstract:Due to the limited space and presence of various equipment within submarine compartments,the variety and concentration of hazardous substances in the air might constantly increase along with the equipment operation and activities or metabolism of operators.Aiming at the problem,this paper proposed an effective detecting method,based on the initial design of a membrane inlet with a buffer cells,that was able to detect p-xylene,toluene and benzene with signal intensity 500,300,160 times higher than that ob tained with glass capillary tubes.The responding time for benzene,toluene and p-xylene was reduced to 6,10 and 15 s,respectively,by measuring the flight mass spectrometry.Meanwhile,several innovative ionization sources including SPI,SPI-CI,and SPI-PEI were developed on the basis of VUV lamps,which further refined the detection limit to 3,4 and 6 ppbv for benzene,toluene and p-xylene,respectively,via a single photon ionization technique.This paper also introduced some progresses in online detection tech nologies of atmospheric pollutants with flight mass spectrometry being the core technique.In short,The de tection device presented here offers a solution for the detection of harmful gas in submarine compartments.