Kinematic Calibration for Articulated Arm Coordinate Measuring Machines Base on Particle Swarm Optimization

Guanbin Gao,Wen Wang,Keng Lin,Zichen Chen
DOI: https://doi.org/10.1109/icicta.2009.54
2009-01-01
Abstract:Kinematic calibration is one of the most important ways to improve the accuracy of articulated arm coordinate measuring machines (AACMMs). The kinematic model of six degree of freedom (6-DOF) AACMMs was established by Denavit-Hartenberg method. One calibration technology base on particle swarm optimization (PSO) was presented, which used single-point repeatability as the fitness function and kinematic parameters as the position vector. An adaptive inertia weight was presented to improve the ability of global exploration and local exploitation of the PSO. A 6-DOF AACMM was used to carry out experimental tests in order to evaluate the efficiency of the self-calibration technology, showing that after calibration the accuracy is improved greatly.