A-362 Hemolysis Detection and Flagging of Potassium Results on GEM Premier 7000 with iQM3

M Butler,L Adib,N McNary,S Balasubramanian,P Pamidi
DOI: https://doi.org/10.1093/clinchem/hvae106.356
IF: 12.114
2024-10-27
Clinical Chemistry
Abstract:BackgroundBlood gas analyzers that report whole blood potassium (K + ) are vulnerable to inaccurate results due to in vitro hemolysis. GEM Premier 7000 with iQM3 (GEM 7000) automatically detects hemolysis (plasma free hemoglobin) and flags impacted K + results. Here, we report the hemolysis impact on K + measurements and the flagging approach used in the GEM 7000 system.MethodsHemolysis interference was assessed according to CLSI EP07c at low (3.8 mmol/L) and high (5.3 mmol/L) K + levels. Ultrasonically-lysed whole blood was used to generate a high test pool (100% = 1000 mg/dL hemoglobin). This was mixed volumetrically with a low pool (0% = native blood) to create additional test pools (25%, 50%, and 75%) and analyzed on three (3) GEM 7000 systems. K + bias at each test pool was calculated from the low pool to generate hemolysis interferograms. Linear regression was performed to determine the hemolysis level that exceeds the interference limit of 7%.ResultsHemolysis exceeding the interference limit was ≥ 116 mg/dL for the low K + and ≥ 147 mg/dL for the high K + level. These correspond to an elevation of 0.27 mmol/L and 0.37 mmol/L of K + , respectively. Based on this study, the GEM 7000 triggers a hemolysis flag when the detected hemolysis exceeds 115 mg/dL. By default, GEM 7000 hemolysis flags are categorized into mild (116 -300 mg/dL), moderate (301 - 400 mg/dL), and gross (≥401 mg/dL) hemolysis.ConclusionsThe magnitude of the observed increase in K + (0.277 - 0.295 mmol/L for every 100 mg/dL of hemoglobin) on the GEM 7000 due to in vitro hemolysis is consistent with previously published reports in the literature. Automated hemolysis detection and flagging of K + results in GEM 7000 overcome the limitation of the inability to detect hemolysis in blood gas samples.
medical laboratory technology
What problem does this paper attempt to address?