Developing a compact coal quality rapid detection integrated machine based on laser‐induced breakdown spectroscopy

Weiye Lu,Qi Yang,Ziyu Yu,Xiangbo Zou,Huaiqing Qin,Gongda Chen,Weizhe Ma,Junling Zhuo,Chuangting Chen,Xiaoxuan Chen,Shunchun Yao,Jidong Lu
DOI: https://doi.org/10.1002/mop.33991
IF: 1.311
2024-01-05
Microwave and Optical Technology Letters
Abstract:Abstract Coal's calorific value and carbon content are crucial in calculating carbon emissions. Accurately detecting these two indicators is of great significance for carbon accounting. In this study, we developed a compact coal quality rapid detection integrated machine based on laser‐induced breakdown spectroscopy (LIBS), which can directly measure coal particle flow. A partial least squares model, based on data set selection according to cluster analysis results, was applied to establish the relationship between coal quality and plasma spectra. The R 2 of the calorific value is .93, the root mean square error of prediction (RMSEP) is 0.41 MJ/kg, and the mean absolute error (MAE) is 0.33 MJ/kg. The R 2 of carbon content is .94, the RMSEP is 0.97%, and the MAE is 0.91%. These results indicated that the developed compact coal quality rapid detection integrated machine could conduct precise coal quality analysis in real time.
engineering, electrical & electronic,optics
What problem does this paper attempt to address?