Magnetic relaxation switching biosensor for one-step detection of Vibrio parahaemolyticus based on click chemistry-mediated sol-gel system

Le Chen,Dongyu Zhang,Fan Yang,Zhongxiu Chen,Yangbo Wu,Jianjun Xie,Zhiyong Guo
DOI: https://doi.org/10.1016/j.snb.2023.134386
2023-08-04
Abstract:In this study, a magnetic relaxation switching (MRS) biosensor was developed for one-step detection of pathogenic bacteria Vibrio parahaemolyticus (VP). The biosensor is based on a click chemistry-mediated sol-gel system and utilizes low-field nuclear magnetic resonance (LF NMR) detection. VP aptamer (Apt)-coated hollow mesoporous silica microspheres (HMSMs) with saturated sodium ascorbate (SAA) solution sealed inside, &Apt, are designed as the signal unit. When the target VP is present, the Apt binds to VP and releases SAA from the signal unit. The released SAA reduces Cu 2+ in the test solution to Cu + , which catalyzes the click chemistry reaction between azide-modified polyethylene glycol (PEG-Azide) and acetylene-modified polyacrylic acid (PAA-Alkyne). The sol-gel transition is triggered, a large amount of "free" water is converted into "bound" water, resulting in the decrease of transverse relaxation time (T 2 ). Under the optimal experimental conditions, the variation of T 2 (ΔT 2 ) was linearly related to the logarithm of VP concentration in the range of 10 ∼ 1.0 × 10 8 CFU/mL, with a limit of detection of 5 CFU/mL. The assay demonstrated good selectivity, stability, and reproducibility. It is completed in one step, holding promising applications in the rapid field detection of pathogenic bacteria.
chemistry, analytical,electrochemistry,instruments & instrumentation
What problem does this paper attempt to address?