Limited Revision with a Newly Designed Hinge Device for the Treatment of Mega‐Prosthesis Hinge Failure: Two Case Reports

Kai Zheng,Xiu‐chun Yu,Ming Xu,Hao‐cheng Cui,Jing‐yu Zhang,Yue Wei
DOI: https://doi.org/10.1111/os.14100
2024-06-05
Orthopaedic Surgery
Abstract:We here present a new technical method, mega‐prostheses partial revision surgery, to repair mega‐prostheses having design defects in hinge devices. Unlike a traditional hinge replacement surgery, this partial revision surgery modifies the hinge mode of the original prosthesis, thereby completely solving the clinical problems. Retaining a stable mega‐prosthesis fixation handle as much as possible for the complete replacement of the hinge device of the mega‐prosthesis is a challenging task. We redesigned and processed a new axial yoke hinge device for two patients and revised their tibial side prosthesis. (A) Original surgical incision was used to fully expose the original mega‐prosthesis. (B) The original hinge device was removed. (C) The original femoral prosthesis was preserved. (D) Tibial side prosthesis and intramedullary bone cement was removed. (E) New tibial side prosthesis and hinge device was installed. Background Surgical treatment for hinge failure in mega‐prosthesis continues to be a challenge. This study introduces a new method for treating hinge failure by using a unilateral prosthesis and hinge revision. Case Presentation We here present two patients who underwent mega‐prosthesis reconstruction after resection of osteosarcoma in the distal femur. To address the issue of knee hyperextension after mega‐prosthesis reconstruction, one patient underwent three revision surgeries, two surgeries were performed using the original hinge, and one surgery involved a newly designed hinge. To resolve the problem of dislocation, one patient underwent three revisions, with the first two revisions not involving hinge replacement and the third revision involving a newly designed hinge. Two replacements of unilateral prosthesis and hinge renovations were successful. Conclusions Unilateral prosthesis and newly designed hinge device revision are effective in treating the failure of old‐fashioned mega‐prosthesis hinges.
orthopedics
What problem does this paper attempt to address?