An integrated mechanical degradation model to explore mechanical response of a bioresorbable polymeric scaffold

A.R. Abaei,Connor J. Shine,T.J. Vaughan,W. Ronan
DOI: https://doi.org/10.1016/j.jmbbm.2024.106419
IF: 4.042
2024-01-26
Journal of the Mechanical Behavior of Biomedical Materials
Abstract:Simulation of bioresorbable medical devices is hindered by the limitations of current material models. Useful simulations require that both the short- and long-term response must be considered; existing models are not physically-based and provide limited insight to guide performance improvements. This study presents an integrated degradation framework which couples a physically-based degradation model, which predicts changes in both crystallinity ( Xc ) and molecular weight ( Mn ), with the results of a micromechanical model, which predicts the effective properties of the semicrystalline polymer. This degradation framework is used to simulate the deployment of a bioresorbable PLLA (Poly (L-lactide) stent into a mock vessel and the subsequent mechanical response during degradation under different diffusion boundary conditions representing neointimal growth. A workflow is established in a commercial finite element code that couples both the immediate and long-term responses. Clinically relevant lumen loss is reported and used to compare different responses and the effect of neo-intimal tissue regrowth post-implantation on degradation and on the mechanical response is assessed. In addition, the effects of possible changes in Xc , which could occur during processing and stent deployment, are explored.
engineering, biomedical,materials science, biomaterials
What problem does this paper attempt to address?