Microbiologically influenced corrosion can cause a dental implant rejection

Eleonora Hubenova,Mario Mitov,Yolina Hubenova
DOI: https://doi.org/10.1016/j.electacta.2024.144087
IF: 6.6
2024-03-20
Electrochimica Acta
Abstract:In the present study, we apply electrochemical and molecular biology approaches to establish whether microbiologically influenced corrosion (MIC) may be the reason for dental implant rejection. Titanium-based implants have been incubated in vitro in Ringer/glucose solution in the presence and absence of a salivary microbiota and the change in their corrosion potential and current density was monitored over time by potentiodynamic polarization. The charge transfers processes occurring on the Ti-based implant surface during the initial bacterial adhesion and after biofilm formation were analyzed by electrochemical impedance spectroscopy. The components of the equivalent electrical circuit as well as the heterogeneous electron transfer rate constants were estimated. In parallel, bacterial species from the salivary microbiota have been identified by sequencing of the 16S ribosomal amplicon method, and their contribution to the corrosion processes of dental implants was verified and compared to that of native saliva in vitro. The scanning electron microscopy observation and energy dispersive spectroscopy analysis of the implant surface after long-term exposure to a bacterial-containing environment revealed more detailed features of the occurring corrosion processes.