Design of a limited-access facility and safety program for a genetic toxicology laboratory

J Inmon,T Vaughan,J Morris
DOI: https://doi.org/10.1080/15298668591394860
Abstract:A limited-access facility has been designed as a result of the need for laboratories for testing hazardous materials found in the environment. The facility design features include room air flow and filtration, hood types, sink design and placement, design of countertops, type of flooring and wall sealant, and traffic flow within the laboratories. These laboratories required the diversity to handle weighing of stock hazardous materials, preparation and handling of aliquots, maintenance of dosed animals as well as sterile conditions required for tissue culture and continuous cell culture methods. A safety and health program was also developed which included specific dress (e.g., scrub suit, TYVEK jumpsuit, gloves, safety glasses and safety shoes), safety advisory group, safety response group, medical monitoring program and training of current and new staff members. The design and use of the facility are continuously reevaluated and changes are made as necessitated by either research needs or improved safety methods.