Cranial ultrasound in preterm infants ≤ 32 weeks gestation—novel insights from the use of very high-frequency (18-5 MHz) transducers: a case series

Francesca Miselli,Isotta Guidotti,Marianna Di Martino,Luca Bedetti,Chiara Minotti,Eugenio Spaggiari,Giovanni Malmusi,Licia Lugli,Lucia Corso,Alberto Berardi
DOI: https://doi.org/10.1007/s00431-024-05627-y
2024-06-05
European Journal of Pediatrics
Abstract:The quality of cranial ultrasound has improved over time, with advancing technology leading to higher resolution, faster image processing, digital display, and back-up. However, some brain lesions may remain difficult to characterize: since higher frequencies result in greater spatial resolution, the use of additional transducers may overcome some of these limitations. The very high-frequency transducers (18-5 MHz) are currently employed for small parts and lung ultrasound. Here we report the first case series comparing the very high-frequency probes (18-5 MHz) with standard micro-convex probes (8-5 MHz) for cranial ultrasound in preterm infants. In this case series, we compared cranial ultrasound images obtained with a micro-convex transducer (8-5 MHz) and those obtained with a very high-frequency (18-5 MHz) linear array transducer in 13 preterm infants ≤ 32 weeks gestation (9 with cerebral abnormalities and 4 with normal findings). Ultrasound examinations using the very high-frequency linear transducer and the standard medium-frequency micro-convex transducer were performed simultaneously. We also compared ultrasound findings with brain MRI images obtained at term corrected age. Ultrasound images obtained with the very high-frequency (18-5 MHz) transducer showed high quality and accuracy. Notably, despite their higher frequency and expected limited penetration capacity, brain size is small enough in preterm infants, so that brain structures are close to the transducer, allowing for complete evaluation.
pediatrics
What problem does this paper attempt to address?