Intensity correction in surface-coil MR imaging.

L. Axel,J. Costantini,J. Listerud
DOI: https://doi.org/10.2214/AJR.148.2.418
1987-02-01
Abstract:The use of surface coils in MR imaging has made it possible to obtain images of superficial structures with improved signal-to-noise ratio relative to conventional circumferential receiver coils [1 -4]. However, the signal intensity distribution in surface-coil images is inherently nonuniform; the decreased sensitivity to signal from more distant regions implies a conresponding decreased sensitivity to noise from those regions. As the total noise detected is uniformly distributed over the image, there is a net increase in the signal-to-noise ratio in the image of regions closer to the coil as compared with the conventional receiver coil [1 , 5]. The resulting range of relative intensity over surface-coil images makes it difficult to display them properly or to analyze them quantitatively. In addition, the local image contrast will decrease proportionally to the average local intensity. We report a method for correcting such surface-coil images so as to produce a uniform relative intensity over the region being imaged.