Heidelberg retina flowmeter parameters at the papilla in healthy subjects.

S. Griesser,A. Lietz,S. Orgl,A. Schtzau,P. Hendrickson,J. Flammer,I. Haefliger
DOI: https://doi.org/10.5301/EJO.2008.5326
European Journal of Ophthalmology
Abstract:PURPOSE To assess the Heidelberg Retina Flowmeter (HRF) parameters "volume", "flow", and "velocity", at the papilla in healthy subjects. METHODS HRF measurements were taken at the papilla (5 x 20), superficially at level of the retina and at the bottom of the excavation. The effect of increasing frame size (1x1 to 50x50 pixels) on HRF values was assessed in ten subjects. HRF parameters were calculated (50x50 pixels) for 150 eyes of 150 subjects. To assess short-term reliability, measurements were repeated five times in ten subjects. RESULTS With 50x50 pixels the location of the frame had no influence on HRF values. Reliability was > 90%. Values were significantly higher (p<0.001) in the superficial than in the deeper papillary layers. The correlation between HRF parameters was good (r2>0.85). CONCLUSIONS A low magnification (5x20) and a 50x50 frame allows a global assessment of HRF parameters at the papilla with high reliability. In healthy eyes, the HRF values are influenced by the level where measurements are made at the papilla. This might be of importance in glaucoma patients with excavated papilla. (Eur J Ophthalmol 1999; 9: 32-6).
What problem does this paper attempt to address?