Homotopy analysis method to study a quadrupole mass filter.

S. S. Chaharborj,S. Kiai,M. R. A. Bakar,I. Ziaeian,Y. Gheisari
DOI: https://doi.org/10.1002/jms.2041
2012-08-01
Journal of Mass Spectrometry
Abstract:The homotopy analysis method (HAM) is applied to study the behavior of a hyperbolic rods of quadrupole mass filter and a sinusoidal potential form V(ac)  cos(Ωt). Numerical computation method of a 20th-order HAM is employed to compare the physical properties of the confined ions with fifth-order Runge-Kutta method. Also, comparison is made for the first stability region, the ion trajectories in real time, the polar plots, and the ion trajectory in x - y plan. The results show that the two methods are fairly similar; therefore, the HAM method has potential application to solve linear and nonlinear equations of the charge particle confinement in quadrupole field.
What problem does this paper attempt to address?