Proof of Concept of a New Revision Procedure for Ceramic Inlays of Acetabular Cups Using a Shape-Memory Alloy Actuator System

Christian Rotsch,Karoline Kemter-Esser,Johanna Dohndorf,Kerstin Funke,Christoph-Eckhard Heyde,Welf-Guntram Drossel
DOI: https://doi.org/10.3390/bioengineering11090868
2024-08-27
Abstract:The revision of ceramic inlays of acetabular cups is a challenging surgical procedure. The mechanical impact during the inlay extraction process can damage the ceramic or metal cup rim. To avoid these risks, a concept for a new revision procedure was developed. It is based on an actuator system, which allows a non-destructive release of the ceramic inlay. To integrate the actuator system, different design concepts of acetabular cup components were investigated, and an actuator based on shape-memory alloy (SMA) wires was developed. The process chain for the actuator, starting from nickel-titanium wires manufactured into the actuator geometry by laser welding and thermo-mechanical treatment for the shape setting process up to the functionality evaluation of the actuator system, was implemented on a laboratory scale. The new revision procedure is based on a phase transformation of the SMA wire actuator, which was obtained through two methods-applying an electrical current by an instrument and rinsing the wire with heated water. The phase transformation of the actuator resulted in a contraction between 3.2% and 4.3% compared to its length after pre-stretching and was able to release the ceramic inlay from the cup. Therefore, the developed actuator design and process chain is a proof of concept towards a new revision procedure for modular acetabular cups.
What problem does this paper attempt to address?