Impact of Needle Design and Suture Gauge on Tissue Tearing During Skin Suturing: A Comparative Analysis

Christian T Potter,McKenzie E Maloney,Alexandria M Riopelle,Gary M Fudem,Carl F Schanbacher
DOI: https://doi.org/10.1097/DSS.0000000000004389
2024-09-26
Abstract:Background: Surgeons face numerous choices in selecting sutures for skin closure, with potential adverse effects such as tissue tearing. Objective: To investigate the influence of needle design and suture gauge on tissue tearing during suturing procedures. Materials and methods: The authors tested the tear-through force in Newtons for 3 needle types and 3 suture gauges using an artificial skin model and a professional-grade tensiometer. Suture material was secured into the skin model, and force was applied to the suture at a constant rate, resulting in tearing. Force-displacement and force-time curves were generated. Evaluation included conventional cutting (PC-3), reverse cutting (PS-3), and taper point (BB) needles with a 5-0 polypropylene suture. In addition, nylon sutures with a reverse cutting needle (PS-2) were tested at 3 suture gauges (5-0, 4-0, 3-0). Results: The mean tear-through forces for PC-3, PS-3, and BB were 3.26 N, 3.75 N, and 4.07 N, respectively. For the 5-0, 4-0, and 3-0 nylon sutures, the mean tear-through forces were 3.44 N, 3.81 N, and 4.04 N, respectively. Statistical analysis revealed a significant impact of suture gauge size (p < .001) and needle geometry (p < .001) on tear-through force. Conclusion: Larger suture diameter and taper needles minimize tissue tearing.