[Application of chromatographic retention time correction in untargeted screening]

Fei Shen,Yifei Fang,Zongli Huo,Feng Zhu,Hao Zhang,Yonglin Zhou
DOI: https://doi.org/10.19813/j.cnki.weishengyanjiu.2024.05.015
Abstract:Objective: Chromatographic retention time correction is one of the important steps to effectively improve the accuracy of identification. This article proposed a strategy for untargeted screening of biological samples based on retention time correction. Methods: A pre-treatment method for biological samples was established. The conditions of liquid chromatography and mass spectrometry were optimized. Fourteen compounds were selected as calibration agents. The retention time correction of different samples, different injection time, different brands of instruments, changing chromatographic column and changing mobile phase were investigated. Results: Calibration agents had a wide coverage, good stability and no interference with sample determination. They could be uniformly distributed in the chromatogram in both positive and negative ion modes. The chromatogram was divided into several time intervals. Calibration agents in each time period were used for retention time linear correction, and the correction effects were good. Conclusion: The retention time correction method could eliminate the retention time drift caused by experimental conditions, improve the accuracy of qualitative analysis, and help to solve the problem of high false positive result based on mass spectrum information.
What problem does this paper attempt to address?