A Reliable and Straightforward Combination of vortex-assisted Dispersive Magnetic solid-phase Extraction with Direct Sorbent Sampling for Highly Sensitive Silver Determination in Water Samples Using Flame and Metal Furnace AAS

Diego Prudencio Dos Santos,Milena do Prado Ferreira,Kristiany Moreira Diniz,Mariana Gava Segatelli,Eduardo Costa de Figueiredo,César Ricardo Teixeira Tarley
DOI: https://doi.org/10.1007/s00128-024-03955-1
2024-09-21
Abstract:In the present paper, the assessment of vortex-assisted dispersive magnetic solid-phase extraction using amino-functionalized mesoporous combined with direct magnetic sorbent sampling (DMSS) in flame or furnace atomic absorption spectrometry (FAAS or FF-AAS) was demonstrated for highly sensitive silver determination in water samples. The developed method showed significant enrichment factors compared to conventional pneumatic nebulization by FAAS, 607 for DMSS-FF-AAS and 114 for DMSS-FAAS. The analytical curve showed linearity in the range from 5.0 to 70.0 µg L- 1 and 1.0 to 15.0 µg L- 1 and limits of detection of 0.59 and 0.09 µg L- 1 for DMSS-FAAS and DMSS-FF-AAS, respectively. The intra and inter-day precision evaluated as a percentage of the relative standard deviation (RSD,%) ranged from 1.89 to 4.71% for levels of 25.0 and 65.0 µg L- 1. The method was applied in different kinds of water samples without matrix effects, yielding recovery values from 90 to 110%.
What problem does this paper attempt to address?