In vivo haemodynamic evaluation of the Hall-Kaster central flow prosthetic heart valve.

B. Semb,A. Aasen,G. Woxholt,F. Resch,A. M. Refsum,T. Frøysaker,S. Nitter‐Hauge,K. Stadsklev,P. Lilleaasen
DOI: https://doi.org/10.3109/00365517909108164
Scandinavian Journal of Clinical and Laboratory Investigation
Abstract:The first in vivo haemodynamic observations of the recently developed Hall-Kaster central flow prosthetic heart valve are reported. Haemodynamic evaluation of valve function was performed both in acute and chronic experiments following implantation of the Hall-Kaster prosthesis in the mitral position in dogs. These studies showed a low transvalvular pressure gradient, an effective opening angle of about 70 degrees and good diastolic flow through both the large and small orifices of the prosthesis. A slow rotational movement was depicted by the disc during opening and closure.