Preliminary Insights Into Early Ambulation and Hemostasis After Atrial Fibrillation Ablation Using the Perclose Vascular Closure Device

Ryuki Chatani,Kota Morikawa,Mitsuru Yoshino,Hiroshi Tasaka,Kazushige Kadota
DOI: https://doi.org/10.1253/circrep.CR-24-0045
2024-08-01
Abstract:Background: The introduction of the Perclose vascular closure device (VCD) for atrial fibrillation (AF) ablation procedures in Japan is expected to facilitate early ambulation. Methods and results: The study population comprised 554 patients undergoing AF ablation, of whom 377 successfully underwent Perclose VCD deployment at the end of the procedure. At 4 h after the procedure, 94% (353/377 patients) achieved early hemostasis and 57% (213/377 patients) achieved early ambulation. Notably, 0.5% (2/377 patients) developed acute deep vein thrombosis. Conclusions: The Perclose VCD demonstrated promising efficacy in achieving early hemostasis, potentially facilitating early ambulation and discharge in appropriate patients.