Design and Modelling of Continuum Robot for Endoscopic Submucosal Dissection Surgery With Lifting Force Estimation

Xingyao Zhang,Fuxin Du,Gang Zhang,Ke Wu,Gang Zheng,Yibin Li,Rui Song
DOI: https://doi.org/10.1002/rcs.2670
Abstract:Background: Endoscopic submucosal dissection (ESD) is an effective treatment for early-stage gastrointestinal cancers. However, traditional surgical instruments lack accuracy and force-sensing. Methods: A new type of continuum robot for ESD is designed. An accurate static model of the proposed continuum robot is established, considering cases where the robot bends into C-shapes and S-shapes. A force estimation method based on an accurate static model is proposed. Then, the accuracy of the static model and force estimation is verified through experiments. Finally, an ex-organ experiment is carried out. Results: The average position error of the proposed static model is 0.72 mm, accounting for 2.57% of the total robot length. The average error of force estimation is 19.53 mN. By gripping and cutting ex-porcine gastric mucosa, the robot's functionality is validated. Conclusion: This paper contributes to precise control and safe interaction of continuum robots.
What problem does this paper attempt to address?