Using jitter analysis with concentric needle electrodes to assess disease status and treatment responses in myasthenia gravis

Vinaya Bhandari,Ajith Sivadasan,Carolina Barnett-Tapia,Hans Katzberg,Vera Bril
DOI: https://doi.org/10.1016/j.cnp.2024.06.004
2024-07-14
Abstract:Objective: This study assesses the utility of jitter analysis with concentric needles to evaluate disease severity in myasthenia gravis (MG), correlate changes in jitter with clinical status as well as identify reasons for any discordance. Methods: We performed a retrospective chart review of 82 MG patients and extracted data on demographics, MG subtype, antibody status, clinical scales, electrophysiology, and interventions at baseline and follow-up. Results: Baseline MGII scores correlated with jitter (r = 0.25, p = 0.024) and abnormal pairs (r = 0.24, p = 0.03). After 28 months, MGII scores correlated with jitter (r = 0.31, p = 0.006), abnormal pairs (r = 0.29, p = 0.009), and pairs with blocks (r = 0.35, p = 0.001). Changes in MGII scores correlated with changes in jitter (r = 0.35, p = 0.002), abnormal pairs (r = 0.27, p = 0.014), and pairs with blocks (r = 0.36, p = 0.001). Conclusions: Concentric needle jitter analysis may have the potential to evaluate baseline and sequential disease severity in MG. Significance: This study highlights the potential for improved MG patient care through precise assessment and management using concentric needle jitter analysis to improve the accuracy of MG diagnosis and monitoring of disease activity.
What problem does this paper attempt to address?