Tissue Radiologic and Pathologic Response to Biphasic Pulsed Electric Field Technology in a Porcine Model

Rosa Kaviani,Robert E Neal,David W Hunter,William S Krimsky
DOI: https://doi.org/10.1016/j.jvir.2024.08.015
2024-08-26
Abstract:Purpose: To evaluate the radiologic, pathologic, and safety characteristics of a commercially available pulsed electric field (PEF) ablation system in a porcine model. Materials and methods: Monopolar biphasic PEF ablation was delivered to the liver, kidney, and longissimus dorsi muscle through a single needle via a percutaneous or open approach with the AliyaTM System. Six animals were ablated and evaluated in two cohorts (Day 3 and Day 28). Muscle, kidney, and liver were ablated in each animal. Intraprocedural Cone Beam CT, follow-up weekly CT, blood serology, gross pathology, and histopathology were performed to characterize the radiographic evolution and tissue response. Results: There were no adverse events, no findings of ECG abnormalities, and serological values returned to baseline by day 28. Ablation zones were visible on non-contrast CT images during follow-up. Most identified zones became radiographically smaller over time, with some fully resolved by day 28. The relative decrease in gross ablation zone volume in the liver and skeletal muscle was 20%, and 26%, respectively, while kidney sites grew by 22%. Ablation sites were focal and were contained within the intended target tissue without extension to non-target tissues or collateral structures. Conclusion: The biphasic PEF system evaluated here resulted in a safe and predictable ablation response, with preservation of structural tissues in an animal model, offering an alternative to thermal ablative modalities, particularly near critical structures.
What problem does this paper attempt to address?