Validation of two LCHRMS methods for large-scale untargeted metabolomics of serum samples: Strategy to establish method fitness-for-purpose

Sietske Grijseels,Terje Vasskog,Pia J Heinsvig,Torbjørn N Myhre,Terkel Hansen,Marie Mardal
DOI: https://doi.org/10.1016/j.chroma.2024.465230
2024-09-13
Abstract:Untargeted metabolomics by LCHRMS is a powerful tool to enhance our knowledge of pathophysiological processes. Whereas validation of a bioanalytical method is customary in most analytical chemistry fields, it is rarely performed for untargeted metabolomics. This study aimed to establish and validate an analytical platform for a long-term, clinical metabolomics study. Sample preparation was performed with an automated liquid handler and four analytical methods were developed and evaluated. The validation study spanned three batches with twelve runs using individual serum samples and various quality control samples. Data was acquired with untargeted acquisition and only metabolites identified at level 1 were evaluated. Validation parameters were set to evaluate key performance metrics relevant for the intended application: reproducibility, repeatability, stability, and identification selectivity, emphasizing dataset intrinsic variance. Concordance of semi-quantitative results between methods was evaluated to identify potential bias. Spearman rank correlation coefficients (rs) were calculated from individual serum samples. Of the four methods tested, two were selected for validation. A total of 47 and 55 metabolites (RPLC-ESI+- and HILIC-ESI--HRMS, respectively) met specified validation criteria. Quality assurance involved system suitability testing, sample release, run release, and batch release. The median repeatability and within-run reproducibility as coefficient of variation% for metabolites that passed validation on RPLC-ESI+- and HILIC-ESI--HRMS were 4.5 and 4.6, and 1.5 and 3.8, respectively. Metabolites that passed validation on RPLC-ESI+-HRMS had a median D-ratio of 1.91, and 89 % showed good signal intensity after ten-fold dilution. The corresponding numbers for metabolites with the HILIC-ESI--HRMS method was 1.45 and 45 %, respectively. The rs median ({range}) for metabolites that passed validation on RPLC-ESI+- was 0.93 (N = 9 {0.69-0.98}) and on HILIC-ESI--HRMS was 0.93 (N = 22 {0.55-1.00}). The validated methods proved fit-for-purpose and the laboratory thus demonstrated its capability to produce reliable results for a large-scale, untargeted metabolomics study. This validation not only bolsters the reliability of the assays but also significantly enhances the impact and credibility of the hypotheses generated from the studies. Therefore, this validation study serves as a benchmark in the documentation of untargeted metabolomics, potentially guiding future endeavors in the field.
What problem does this paper attempt to address?