Bottle-in-bottle reaction device: Portable gas pressure meter for rapid and on-site analysis of oxalate in spinach and tea beverages

Jiayuan Tang,Dali Zhuo,Xiaoyu Dong,Jing Xiao,Shu Zhang,Xiaoli Xiong,Chao Tan,Zhirong Zou
DOI: https://doi.org/10.1016/j.fochx.2024.101638
2024-07-11
Abstract:A gas pressure meter-based portable/miniaturized analytical kit was established for rapid and on-site detection of oxalate. Potassium permanganate (KMnO4) and oxalate solution were mixed together in bottle-in-bottle reaction device, a simple oxidation reaction process occurred within 6 min and carbon dioxide (CO2) was generated, inducing the pressure of the sealed bottle changed, which was measured by a portable gas pressure meter. A detectable range of 0.1-6 μmol mL-1 and a detection limit of 0.064 μmol mL-1 were achieved. The proposed analytical method was further used for the analysis of several real samples (spinach, beverages and water samples), with the recoveries of 89-111%. Considering the interferences from the complicated matrix, calcium chloride (CaCl2) was served as a precipitant, oxalate (C2O4 2-) was precipitated with Ca2+ to form precipitation (CaC2O4), CaC2O4 was then separated from the matrix by centrifuge/filter, eliminating the interferences. It is a rapid, easy-used and interference-free analytical system/device for oxalate on-site and real time analysis.